Gelişmiş Arama

Basit öğe kaydını göster

dc.contributor.authorBasat, H. Cagda
dc.contributor.authorKirkayak, Levent
dc.date.accessioned2019-11-24T21:01:14Z
dc.date.available2019-11-24T21:01:14Z
dc.date.issued2018
dc.identifier.issn2157-9083
dc.identifier.issn2157-9091
dc.identifier.urihttps://dx.doi.org/10.1166/jbt.2018.1906
dc.identifier.urihttps://hdl.handle.net/20.500.12513/3504
dc.descriptionWOS: 000458828200002en_US
dc.description.abstractThere have been numerous methods used for improving glenoid base plate stability in reverse shoulder arthroplasty. In this regard, screw insertion angle is the most significant parameter. Our claim is that screw insertion angle provokes changes in contact stress distribution on humeral spacer. We analyzed effect of 20 different models on humeral spacer with finite element analysis. This was comprised of 5 different screw angles, 4 different abduction positions of shoulder and five intact muscles (deltoid, teres major, infraspinatus, teres minor, subscapularis). The screw insertion angle affected stress distribution of humeral spacer and became more significant as the shoulder abduction reached 90 degrees where the maximum moment to the shoulder arm was applied. In addition, it is shown that screw insertion angle is not only an important parameter for the stress distribution but also the magnitude of the contact stress on humeral spacer due to the changes of direction of the transmitted load during the abduction. Contact stresses on quadrants of the humeral spacer could be influenced by changing screw insertion angle and shoulder abduction angle. Also, a screw insertion angle placement at 17 degrees provided the optimal stress distribution on the humeral spacer at all abduction positions of shoulder.en_US
dc.language.isoengen_US
dc.publisherAMER SCIENTIFIC PUBLISHERSen_US
dc.relation.isversionof10.1166/jbt.2018.1906en_US
dc.rightsinfo:eu-repo/semantics/closedAccessen_US
dc.subjectScrew Insertion Angleen_US
dc.subjectReverse Shoulder Arthroplastyen_US
dc.subjectContact Stressen_US
dc.subjectHumeral Spaceren_US
dc.subjectGlenoid Baseplateen_US
dc.subjectFinite Element Methoden_US
dc.titleStress Effect of Screw Insertion Angle for Base Plate Fixation on Humeral Spacer in Reverse Shoulder Arthroplastyen_US
dc.typearticleen_US
dc.relation.journalJOURNAL OF BIOMATERIALS AND TISSUE ENGINEERINGen_US
dc.contributor.departmentKırşehir Ahi Evran Üniversitesi, Tıp Fakültesi, Cerrahi Tıp Bilimleri, Ortopedi ve Travmatoloji ABDen_US
dc.identifier.volume8en_US
dc.identifier.issue11en_US
dc.identifier.startpage1535en_US
dc.identifier.endpage1542en_US
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanıen_US